Introduction
This Application Note is the second in a series relating to the eQCM 10M™ and assumes that
you have read and understand the material covered in the Application Note, Basics of a
Quartz Crystal Microbalance.
The purpose of this Application Note is to show the reader a simple experiment where we
reduce Cu2+onto an Au electrode and then oxidize Cu back off. We will then use the result to
calculate a calibration factor for the crystal. Conversely, we will show you how to calculate the
molar mass of the species deposited using a calibration factor.
Each crystal has a theoretical calibration factor, however, under typical experimental
conditions, these calibration factors will vary slightly. We will use Cu2+ to calculate a
calibration factor which could then be used in subsequent experiments on the same crystal.
Cu2+ is reduced in a two electron reaction as shown below:

CuSO4(aq) + 2 e- →Cu(s) + SO42-(aq)
Our goal is to reduce Cu2+ onto an Au electrode, then use the frequency decrease and the
charge passed to calculate a calibration factor.

美国Gamry电化学(merbclass.com)热门推荐:综合电化学工作站,多通道电化学工作站,多通道多功能电化学工作站,电化学工作站分辨率,Gamry电化学工作站
香港六合彩公司merbclass.com版权所有 总访问量:39508 地址:上海市杨浦区逸仙路25号同济晶度310室 邮编:200437
电话:86-021-65686006 传真:86-021-65688389 手机: 联系人:汤小姐 邮箱:ytang@gamry.com
GoogleSitemap 技术支持:中国化工仪器网 管理登陆 沪ICP备15019588号-2
联系人
在线客服
用心服务 成就你我

中国化工仪器网

顶级会员

顶级会员

2

推荐收藏该企业网站
友情链接:香港六合彩公司  香港六合彩公司  香港六合彩公司  香港六合彩公司  香港六合彩公司  香港六合彩公司  香港六合彩公司  

免责声明: 本站资料及图片来源互联网文章,本网不承担任何由内容信息所引起的争议和法律责任。所有作品版权归原创作者所有,与本站立场无关,如用户分享不慎侵犯了您的权益,请联系我们告知,我们将做删除处理!